SW_EFIxLPI_1B

Electric field instrument (EFI) Langmuir probe (LP) measurements at 1Hz

Thematic areas: Plasma measurements, Space Weather

Applicable missions: Swarm

Applicable spacecraft: Swarm-A, Swarm-B, Swarm-C

Description

The EFIXLPI_1B Product contains the EFIX_LP_1B plasma data, which are data from the Langimur Probe (LP) of the Electrical Field Instrument (EFI), interpolated at full UTC seconds, the same as the low resolution magnetic data MAG_LR.

Data access

FAST processing

N/A

Preview image

(delivered separately)

File contents

Variable Units Description Dim Type
Timestamp UTC Time stamp 1 CDF_EPOCH
SyncStatus - Synchronization status 1 CDF_UINT2
Latitude deg Geocentric latitude 1 CDF_DOUBLE
Longitude deg Geocentric longitude 1 CDF_DOUBLE
Radius m Distance from the Earth’s center 1 CDF_DOUBLE
U_orbit m/s Spacecraft velocity in the ITRF 1 CDF_DOUBLE
Ne cm-3 Plasma density (electron) 1 CDF_DOUBLE
Ne_error cm-3 Error estimate of plasma density (ne) 1 CDF_DOUBLE
Te K Plasma electron temperature 1 CDF_DOUBLE
Te_error K Error estimate of plasma electron temperature (T_elec) 1 CDF_DOUBLE
Vs V Spacecraft potential 1 CDF_DOUBLE
Vs_error V Error estimate of spacecraft potential (U_pol) 1 CDF_DOUBLE
Flags_LP - Flags indicating the source/method of LP measurements (Ne, Te, Vs) 1 CDF_UINT1
Flags_Ne - Flags characterizing the plasma density measurement 1 CDF_UINT1
Flags_Te - Flags characterizing the electron temperature measurement 1 CDF_UINT1
Flags_Vs - Flags characterizing the spacecraft potential measurement 1 CDF_UINT1

More details

   
Flag
   
Value    
Description    
   
Flags_LP
   
   
   
1    
   
High gain   probe has no errors    
   
3    
   
High gain   probe had errors, instead the low gain probe had to be used. See Flags_Te for   implications.    
   
7    
   
Only one   original data point was used due to sweep or error with a time    
   
Flags_Ne
   
   
   
10    
   
Nominal data,   calibration error for this sample is computed    
   
19    
   
Nominal data,   but calibration error is not computed/out of range for at least one of the   original data points.    
   
20    
   
Nominal data,   calibration error is not computed for both of the original data points.    
   
30    
   
The estimate   for one of the original data points is from the low gain probe. Ne estimate probably has a large random   error    
   
31    
   
The estimate   for both data points is from the low gain probe. Ne estimate probably has a very large   random error    
   
40    
   
Negative   density (high positive values, even extremes, are not flagged)    
   
Flags_Te
   
   
   
10    
   
Nominal data,   calibration error for this sample is computed    
   
19    
   
Nominal data,   calibration error for at least one of the data points is not computed/out of   range    
   
20    
   
Nominal data,   error for both data points is not computed    
   
12    
   
Calibration   error for this sample is computed, but at least one data point had ADC   overflow at the linear bias, high gain probe, tracking ok, discard Te is recommended    
   
15    
   
Calibration   error for this sample is computed, but at least one data point had ADC   overflow at the linear bias, low gain probe, tracking ok, discard Te is recommended    
   
22    
   
Error is not   computed, at least one data point had ADC overflow at the linear bias, high   gain probe, tracking ok, discard Te is   recommended    
   
25    
   
Error is not   computed, at least one data point had ADC overflow at the linear bias, low   gain probe, tracking ok, discard Te is   recommended    
   
32    
   
Failed   tracking, both data points, high gain probe, discard Te is recommended    
   
35    
   
Failed   tracking, both data points, low gain probe, discard Te is recommended    
   
36    
   
Extreme   value, both data points, discard Te is   recommended    
   
40    
   
Negative Te, both data points, discard    
   
Flags_Vs
   
   
   
10    
   
Nominal data,   calibration error for this sample is computed    
   
20    
   
Nominal data,   error for both data points is not computed    
   
25    
   
ADC overflow   at the retarded bias, both data points, discarding Vs is recommend    
   
26    
   
ADC overflow   at the linear bias, both data points, discarding Vs is recommend    
   
32    
   
Failed   tracking, both data points, discarding Vs is recommend    
   
33    
   
Value   unreasonable, both data points, discarding is recommended    

Related resources

Level 1B Product Descriptions:

Level 1B Product Format page:

Changelog

N/A